Long story short, I bought a house with my wife and have been renovating it into the lab of my dreams. When complete, it will have a dedicated wet chemistry bench, 400 square feet of ESD-floored lab space, 200 square feet of office/conference room space, over a mile of multimode fiber and copper Ethernet cabling, and lots of other goodies. There will be a post (or more likely a series) on the setup when it's done, but that's still a few months out.
Among the new toys I've acquired is a LeCroy WaveRunner 8104 oscilloscope - a beast of a scope with four 1 GHz channels at up to 20 Gsps.
New scope is on the right rack (left side scope is an older WaveSurfer 3034) |
There's just one problem... probes.
The stock passive probes have a rated bandwidth of 500 MHz and, as with all passive R-C probes, have severe input impedance roll-off at higher frequencies. According to the datasheet, at ~150 MHz the input impedance drops below 100Ω and it only gets worse from there! As you might imagine, this isn't going to be healthy for signals on a 50Ω line.
LeCroy does sell active probes (such as the ZS1500) but they are not cheap - even secondhand, you're looking at a low four-figure price tag per channel. I might pick one up for comparison purposes eventually, but won't be evaluating them during this post.
In order to explore different probe designs, I built a characterization board on OSHPark's 4-layer process. It has an edge-launch SMA at either end of two 50Ω microstrip traces, a ground plane on layer 2, then the bottom layers are unused. (I could have done this as a 2-layer board, but the traces would have been massive given the increased distance to the ground plane.)
Probe characterization board. The little gold points and the clip are grounds. |
I didn't even bother testing the standard alligator-clip ground wires on the PP022 probes, because I knew these were way too inductive to be of any value whatsoever for high-speed signals. Plus, to make it a fair comparison I needed to give the stock probe a chance to put its best foot forward. All of this testing was done with the spring-tip ground.
The first test signal was a 1 GHz pure sine source built around a Crystek CCSO-914X-1000 oscillator with some additional filtering on the output to remove harmonics.
Test setup. The helping-hand is sitting on top of the coax leading to my scope so it doesn't flail around. |
Closeup of the probe tip |
SMA feedthrough (top) vs PP022 (bottom), same scale. I probably should have turned persistence on to smooth out the traces, but the amplitude difference is strikingly obvious. |
The other probe being tested today was a custom 20:1 transmission line probe. I had high hopes for this design as I had used it with good results on slower circuits in the past, but this was my first time trying it on such fast signals.
Transmission line probe prototype |
The test setup was pretty much the same as before, except I used the helping hands to hold the probe. (I have a 3D printed shell on order that fits the same Pico Tech probe positioners I used on the PP020, but it's not here yet.)
Experimental setup |
At first glance, performance was quite encouraging. The loading was noticeably less (feed-through signal was 2003 mV, vs 2192 un-loaded and 1930 with the stock probe) and the measured signal amplitude was 946 mV, very close to the actual value. This suggests the experimental probe design has a -3 dB bandwidth right around 1 GHz, or double the bandwidth of the stock probe!
Feedthrough signal (top), prototype probe (bottom) |
The experimental setup was identical to before, except that both legs of the test board were used since the GTP transceiver has a differential output. Although I was only probing one leg, I wanted to run clock recovery off the differential signal to minimize jitter in my measurements.
The positive leg of the feed-through differential pair was plotted in the top left quadrant of the scope view, while the probe's output was plotted in the top right. An offscreen filter subtracted the two differential pair legs, then fed the differential signal to a CDR PLL. This same recovered clock was used as the clock source for eye patterns on both signals, displayed in the bottom quadrants below their corresponding inputs.
The PP020 performed surprisingly well.
SMA feedthrough (left), PP020 (right) |
This failure was entirely expected - but what I did not expect was how poorly my own probe design performed on the same test signal. While it was less intrusive to the pass-through signal, the eye measured through the probe was practically nonexistent.
SMA feedthrough (left), my probe (right) |
With no probe, the signal looked quite nice. There was a small amount of ringing on the rising edge, but this didn't impair readability of the signal at all. I also tried another color ramp (matplotlib Viridis) for the eye patterns in hopes of seeing more detail.
Differential signal with no probe |
My probe at 125.25 Mbps |
The obvious hypothesis was that the 38Ω area was an impedance mismatch at either the probe needle or the MMCX connector. I discounted the needle as the mismatch initially, because an 0.51mm diameter needle shank 2mm from a ground lead has an impedance of 244Ω which didn't match the observed data whatsoever.
The next obvious target was the MMCX. I calculated 26Ω impedance for the 1.1mm wide center contact over the ground plane (I had somehow forgotten to put a ground cutout under it). After milling out a cavity, though, performance didn't seem to improve by any detectable amount.
Milled cavity under center pad |
The other possibility is that the lack of a cutout under the needle is problematic. I initially thought this would be OK because the needle is about the same diameter (0.51mm vs 0.41mm) as the PCB trace, but later calculations for an 0.51mm wire above a ground plane using the OSHPark stackup suggest that the impedance of the needle mounting could be as low as 30-40Ω which is consistent with the observed impedance hump.
Finally, the 0.41mm microstrip appears to have been over-etched closer to 0.32mm. If the needle mounting is the impedance mismatch, this would be consistent with the observed 55Ω hump after it. I might try thickening the microstrip with solder to see if this helps bring impedance up closer to 50.